____________________ > Atlanta
> System components
> New Style
> Advanced micro osmometer model 3300 msrp $9385
Advanced micro osmometer model 3300 msrp $9385
ADVANCED MICRO OSMOMETER MODEL 3300 MSRP $9385
Condition: Cosmetically Good Condition. Pre-Owned, Missing Temperature Stick.
The Advanced Instruments Model 3300 Micro-Osmometer determines the osmolality of solutions using freezing point depression (FPD). It is ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications.
Automated, single sample, designed to process a 20 L sample, and one-minute test time, allows the Model 3300 to provide the best accuracy and precision in the market today.
* 20 L is ideal for applications where the sample quantity is limited.
* On-board statistical analysis
Method Freezing Point Depression Osmometry
Osmometry measures the total molar concentration of dissolved solids in any solution. The freezing point of a solution is depressed in direct relation to the amount of solute in solution.
Uses and Applications Emergency Medicine, Internal Medicine/Endocrinology, Urology/Renal Function, Surgery/Critical Care, Obstetrics/Gynecology, Pharmacy, General Quality Control.
Power Requirement: 100 to 130 VAC (50-60 Hz) or 200 to 250 VAC (50/60 Hz)
Fuses (2): 250V time delay (Type T): 1.25-Amp (100-130 VAC); 1-Amp (200-250 VAC)
Memory Backup: Integral lithium cell; 10-year life (typical); (not user-replaceable)
Sample Capacity: Single sample
Readout: 2 line by 24-character vacuum fluorescent display
Communications: DTE EIA-232/V.24 (RS-232); serial port, parallel printer port and optional barcode scanner
Performance at Reference Conditions
Linearity: Less than 1% from a straight line between 0 and 850 mOsm/kg H2O; less than 1.5% above 850 mOsm/kg H2O
Repeatability: 0 to 400 mOsm: 2 mOsm/kg H2O (1 S.D.) 400 to 2000 mOsm: 0.5% (1 S.D.)
Drift: Less than 1 mOsm/kg H2O per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/kg H2O for every 5C (9F) ambient temperature change
Operating Conditions Temperature: 18C to 35C (64F to 95F)
Humidity: 5 to 80% relative humidity (non-condensing)
Storage Temp.: -40C to +45C (-40F to +113)
Start-up Time: 30 seconds from power-on